FDA Device recall Z-1104-2020
Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version 1.4.4 GUDID Register…
- FDA Device
- Class II
- terminated
- Published 2020-02-19
On 2020-02-19 the FDA classified a Class II recall of Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version 1.4.4 GUDID Register… by Dexcom, citing it was reported that the user's low alarm feature on the iOS application were not properly alerting users when the user has enabled the Alert Schedule feature more than 30 days af….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1104-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-11-19 |
| Published | 2020-02-19 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Dexcom |
| Distribution | n/a |
Product
Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804
Reason
It was reported that the user's low alarm feature on the iOS application were not properly alerting users when the user has enabled the Alert Schedule feature more than 30 days after installing the iOS application.
Identifiers
| code_info | Software Number: SW11677 Software Version: 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1104-2020%22