FDA Device recall Z-1101-2013

MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System

On 2013-04-17 the FDA classified a Class II recall of MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System by Toshiba American Medical Systems, citing it was found that there is a problem with the LCD monitor used in the console. The LCD monitor may become blank during use. Specifically, the backlight of the LCD monitor may turn….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1101-2013
ClassificationClass II
Statusterminated
Initiated2012-10-15
Published2013-04-17
Last updated2026-09-13 11:53 UTC
CompanyToshiba American Medical Systems
DistributionUS

Product

MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System.

Reason

It was found that there is a problem with the LCD monitor used in the console. The LCD monitor may become blank during use. Specifically, the backlight of the LCD monitor may turn off while the system is in use, causing the LCD monitor to become blank. However, the other parts of the system continue to operate normally, and the risk of a serious health hazard is therefore very low. The LCD Moni

Identifiers

code_infoProduct Numbers: MRT-1504/A4, MRT-1504/S4, MRT-3010/A5, MZKT-UPKHE/S1. Serial Numbers unique to each consignee.