RecallRadar

FDA Device recall Z-1100-2025

The MSOT Acuity Echo is a Class 4 medical laser device

On 2025-02-19 the FDA classified a Class II recall of The MSOT Acuity Echo is a Class 4 medical laser device by Ithera Medical, citing the MSOT Acuity Echo does not include an adequate instruction for safe operation of the device. The previous version of the user manuals contained a wrong value for the Nominal Oc….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1100-2025
ClassificationClass II
Statusongoing
Initiated2025-01-08
Published2025-02-19
Last updated2026-09-13 11:57 UTC
CompanyIthera Medical
DistributionUS

Product

The MSOT Acuity Echo is a Class 4 medical laser device. The MSOT Acuity is a mobile assembly of electrically-powered components intended for imaging and analysis of soft-tissue and vasculature using combined optoacoustic (photoacoustic) and ultrasound (US) imaging.

Reason

The MSOT Acuity Echo does not include an adequate instruction for safe operation of the device. The previous version of the user manuals contained a wrong value for the Nominal Ocular Hazard Distance (NOHD).

Identifiers

code_info
Model numbers MSOT Acuity Echo (Research System) MSOT Acuity Echo CE MSOT Acuity Echo CE2 UDI's: (01)04262380070004 (0…Model numbers MSOT Acuity Echo (Research System) MSOT Acuity Echo CE MSOT Acuity Echo CE2 UDI's: (01)04262380070004 (01)04262380070004(21)22102 (01)04262380070004(21)22203 (01)04262380070004(21)21901 (01)04262380070004(21)22101 (01)04262380070004(21)22104 (01)04262380070004(21)22105 (01)04262380070004(21)22106 (01)04262380070004(21)22107 (01)04262380070004(21)22201 (01)04262380070004(21)22202 (01)04262380070004(21)22204 (01)04262380070004(21)22205 (01)04262380070004(21)22206 (01)04262380070004(21)22301 (01)04262380070004(21)22401

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1100-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.