RecallRadar

FDA Device recall Z-1100-2018

Malecot Nephrostomy Catheter Set, Product Number: M0064101000, GTIN: 08714729005261 The Malecot Nephrostomy Catheter Set is intended to est…

On 2018-03-28 the FDA classified a Class II recall of Malecot Nephrostomy Catheter Set, Product Number: M0064101000, GTIN: 08714729005261 The Malecot Nephrostomy Catheter Set is intended to est… by Boston Scientific, citing certain Malecot Nephrostomy Catheters are breaking at the mid-shaft bond during use.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1100-2018
ClassificationClass II
Statusterminated
Initiated2017-12-12
Published2018-03-28
Last updated2026-09-13 11:54 UTC
CompanyBoston Scientific
DistributionUS

Product

Malecot Nephrostomy Catheter Set, Product Number: M0064101000, GTIN: 08714729005261 The Malecot Nephrostomy Catheter Set is intended to establish percutaneous nephrostomy drainage and is indicated for the following: Diagnostic Indications: " Antegrade pyelography " Pressure/perfusion study (Whitaker test) Therapeutic Indications: " Nephrostomy catheter drainage " Perfusion chemolysis of renal stones " Percutaneous nephrolithotomy " Percutaneous resection and coagulation of urothelial tumors

Reason

Certain Malecot Nephrostomy Catheters are breaking at the mid-shaft bond during use.

Identifiers

code_info
Expiration Date Range: 1/22/2018 - 10/24/2021; Batch: 16682515 18351959 19879547 16954904 18518065 20221526 17023624…Expiration Date Range: 1/22/2018 - 10/24/2021; Batch: 16682515 18351959 19879547 16954904 18518065 20221526 17023624 18662849 20491742 17174677 18689958 20593965 17547901 18773158 20703428 17634224 18821699 20798225 17797370 18898592 20824604 17881031 19103570 20872724 17990643 19134817 20995561 18006579 19185066 21055681 18141765 19235546 21101919 18280356 19368679 21179390

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1100-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.