FDA Device recall Z-1099-2025

Belmont Medical 3-Spike Disposable Set used with the Belmont Rapid Infuser RI-2. Designed to be used in general operation in hospital or al…

On 2025-02-12 the FDA classified a Class II recall of Belmont Medical 3-Spike Disposable Set used with the Belmont Rapid Infuser RI-2. Designed to be used in general operation in hospital or al… by Belmont Instrument, citing disposable set may leak during priming from a female quick connector due to a crack in the female quick connector, may lead to a potential delay in treatment since the set will ne….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1099-2025
ClassificationClass II
Statusongoing
Initiated2025-01-02
Published2025-02-12
Last updated2026-09-13 11:57 UTC
CompanyBelmont Instrument
DistributionUS

Product

Belmont Medical 3-Spike Disposable Set used with the Belmont Rapid Infuser RI-2. Designed to be used in general operation in hospital or alternate care environments to provide warmed blood and fluids to any patients e10 kg requiring warmed infusion. Part Number:903-00006

Reason

Disposable set may leak during priming from a female quick connector due to a crack in the female quick connector, may lead to a potential delay in treatment since the set will need to be replaced by another set

Identifiers

code_infoUDI-DI: 00896128002022 Lot Numbers: 20240910, 20241002, 20241006, 20241014

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.