FDA Device recall Z-1095-2019

icumedical Oncology Kit w/12" (30cm) Ext Set w/Spiros w/Red Cap, Clamp, Graduated Adapter

On 2019-04-24 the FDA classified a Class I recall of icumedical Oncology Kit w/12" (30cm) Ext Set w/Spiros w/Red Cap, Clamp, Graduated Adapter by Icu Medical, citing there is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the devi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1095-2019
ClassificationClass I
Statusterminated
Initiated2019-02-21
Published2019-04-24
Last updated2026-09-13 11:55 UTC
CompanyIcu Medical
DistributionUS

Product

icumedical Oncology Kit w/12" (30cm) Ext Set w/Spiros w/Red Cap, Clamp, Graduated Adapter; ChemoClave Vented Vial Spike, 20mm; ChemoClave Vented Vial Spike, 13mm; Spiros wRed Cap, REF 3944. Used in IV administration sets. The ChemoCLAVE Cytotoxic Medication Preparation and Delivery System consists of 6 previously cleared components (CLAVE, Spikes, SPIROS, GENIE, Vial Access, and Admin Sets) that can be combined into various configurations intended for use in the preparation and patient administration of cytotoxic medications.

Reason

There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.

Identifiers

code_infoLot Numbers: 3846116, 3855565

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.