Maintenance Kit PH2, Cod. 6997274 The Phoenix Hemodialysis delivery system is intended to be used to provide high flux and low flux hemodia…
- FDA Device
- Class II
- terminated
- Published 2017-02-01
On 2017-02-01 the FDA classified a Class II recall of Maintenance Kit PH2, Cod. 6997274 The Phoenix Hemodialysis delivery system is intended to be used to provide high flux and low flux hemodia… by Baxter Healthcare, citing baxter Healthcare Corporation is issuing a voluntary product recall for several lots of Phoenix Dialysis Machine Preventive Maintenance Kits due to incorrect silicone straight con….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1095-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-01-04 |
| Published | 2017-02-01 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Baxter Healthcare |
| Distribution | US |
Product
Maintenance Kit PH2, Cod. 6997274 The Phoenix Hemodialysis delivery system is intended to be used to provide high flux and low flux hemodialysis, hemofiltration and ultrafiltration on patients weighing 15 Kilograms or more. The Phoenix system is to be used with either high or low permeability dialyzers. The device is intended to be used by trained operators when prescribed by a physician, in a chronic care dialysis facility or acute care unit
Reason
Baxter Healthcare Corporation is issuing a voluntary product recall for several lots of Phoenix Dialysis Machine Preventive Maintenance Kits due to incorrect silicone straight connectors components packaged in these kits. During Preventive Maintenance, the two silicone connectors must be installed in the Phoenix mass-balance section of hydraulic circuit only if the Phoenix machine has already been modified with "CONN. IMPROVING RETR. KIT" (Code SP00721). Installation of these connectors in machines modified with SP00721 may result in fluid leaks during dialysis treatment. These Preventive Maintenance Kits include the FILTER sub-group, which contains the silicone straight connectors that are larger than expected.
Identifiers
| code_info | Product code: 6997274; Lot codes: 160718E, 160720E, 160825E |
|---|