RecallRadar

FDA Device recall Z-1092-2026

Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combi…

On 2026-01-21 the FDA classified a Class II recall of Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combi… by C Rad Positioning Ab, citing due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1092-2026
ClassificationClass II
Statusongoing
Initiated2025-10-09
Published2026-01-21
Last updated2026-09-13 11:58 UTC
CompanyC Rad Positioning Ab
DistributionUS

Product

Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable

Reason

Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.

Identifiers

code_info
Lot Code: Note: The error is not related to any serial numbers or LOT numbers. The error is limited to the software mod…Lot Code: Note: The error is not related to any serial numbers or LOT numbers. The error is limited to the software module cSRS in combination with cMotion module