RecallRadar

FDA Device recall Z-1085-2026

Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B

On 2026-01-21 the FDA classified a Class II recall of Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B by Medline Industries, citing medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1085-2026
ClassificationClass II
Statusongoing
Initiated2025-12-16
Published2026-01-21
Last updated2026-09-13 11:58 UTC
CompanyMedline Industries
DistributionUS

Product

Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359

Reason

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Identifiers

code_info1) UDI/DI 10198459269196 (EA) 40198459269197 (CS), Lot Number 25EBD732; 2) UDI/DI 10198459326318 (EA) 40198459326319 (CS), Lot Number 25FBA103.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1085-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.