AlluraXperFD20/10 System Code: (1) 722029

On 2025-02-12 the FDA classified a Class II recall of AlluraXperFD20/10 System Code: (1) 722029 by Philips Medical Systems Nederland B V, citing potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the pa….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1084-2025
ClassificationClass II
Statusongoing
Initiated2025-01-13
Published2025-02-12
Last updated2026-09-13 11:57 UTC
CompanyPhilips Medical Systems Nederland B V
DistributionUS

Product

AlluraXperFD20/10 System Code: (1) 722029

Reason

Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

Identifiers

code_infoUDI: 00884838054219(21)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.