RecallRadar

FDA Device recall Z-1084-2015

The MAYFIELD¿ Ultra 360 Base Unit for head positioner, one unit per package

On 2015-02-18 the FDA classified a Class II recall of The MAYFIELD¿ Ultra 360 Base Unit for head positioner, one unit per package by Integra Lifesciences, citing through an investigation of complaints, Integra found that there is the possibility for the locking handle mechanism of the Base Units to fail during use.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1084-2015
ClassificationClass II
Statusterminated
Initiated2014-12-19
Published2015-02-18
Last updated2026-09-13 11:53 UTC
CompanyIntegra Lifesciences
DistributionUS

Product

The MAYFIELD¿ Ultra 360 Base Unit for head positioner, one unit per package. Product Usage: This product provides the ability to position and hold a patient head while the patient is in the prone, supine, lateral or park-bench and sitting positions. It is intended to be used during diagnostic examinations or surgical procedures where a rigid support between the surgical table and a headrest or skull clamp is necessary and positional freedom is required.

Reason

Through an investigation of complaints, Integra found that there is the possibility for the locking handle mechanism of the Base Units to fail during use.

Identifiers

code_info
All Lot numbers of Mayfield Ultra 360 Base Units, Catalogue #'s A2009 and A2029. Affected Lot Numbers For Catalogue…All Lot numbers of Mayfield Ultra 360 Base Units, Catalogue #'s A2009 and A2029. Affected Lot Numbers For Catalogue No. A2009 73149, 73355, 73356, 74396, 74730, 74398, 74399, 75796, 75816, 74523, 75838, 75859, 74524, 75873, 74890, 75939, 74892, 76420, 74893, 74894, 74895, 76156, 77225, 76157, 76158, 77831, 76159, 77957, 76160, 78012, 76161, 78548, 77292, 78666, 79067, 79068, 79069, 79071, 80013, 79073, 81782, 79074, 81918, 79075, 79748, 80094, 82194, 80095, 80668, 80669, 84171, 80671, 84288, 82736, 82737, 82738, 85896, 82739, 86400, 82740, 82741, 82742, 84732, 84733, 84734, 84735, 85693, 86817, 86818, 86819, 86820, 86821, 86822, 89453, 89455, 89456, 89457, 92296, 92297, 92298, 92299, 92300, 92301, 94653, 94654, 94655, 99798, 98558, 98559, 98560, 98561, 98562, 94658, 103127, 103128, 103129, 103130, 103131, 104890, 114978, 115809, E19251, 123051 Affected Lot Numbers For Catalogue No. A2029 84180, 84487, 85519, 85815, 85631, 84887, 84888, 86533, 86534, 84889, 86828, 86829, 89302

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1084-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.