FDA Device recall Z-1083-2024
MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007
- FDA Device
- Class II
- ongoing
- Published 2024-02-21
On 2024-02-21 the FDA classified a Class II recall of MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007 by Sedecal, citing while cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1083-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-10-14 |
| Published | 2024-02-21 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Sedecal |
| Distribution | US |
Product
MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007
Reason
While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.
Identifiers
| code_info | Model 712001 - UDI-DI 00884838028586, all serial numbers; Model 712002 - UDI-DI 00884838082373, all serial numbers; M…Model 712001 - UDI-DI 00884838028586, all serial numbers; Model 712002 - UDI-DI 00884838082373, all serial numbers; Model 712004 - UDI-DI 08436046002173, all serial numbers; Model 712006 - UDI-DI 00884838091504, serial numbers: SN20410556 and SN21410184; Model 712007 - UDI-DI)08436046002357, serial numbers: 26, 27, 19410038, 19410044, 19410046, 21410261, 21410266, 21410280, 21410288, 21410290, 21410291, 21410293, 21410294, 21410339, 21410340, 21410341, 21410342, 21410345, 22410022, 22410023, SN0410157, SN1410049, SN19410034, SN19410035, SN19410036, SN19410037, SN19410071, SN19410092, SN19410093, SN19410094, SN19410097, SN19410098, SN19410099, SN19410100, SN19410101, SN19410102, SN19410115, SN19410116, SN19410123, SN19410127, SN19410129, SN19410135, SN19410136, SN19410137, SN20410029, SN20410030, SN20410032, SN20410033, SN20410034, SN20410044, SN20410045, SN20410049, SN20410060, SN20410063, SN20410064, SN20410065, SN20410066, SN20410071, SN20410072, SN20410073, SN20410074, SN204 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1083-2024%22