FDA Device recall Z-1082-2019
Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine & Traum…
- FDA Device
- Class I
- terminated
- Published 2019-04-24
On 2019-04-24 the FDA classified a Class I recall of Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine & Traum… by Brainlab Ag, citing in certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1082-2019 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2019-02-27 |
| Published | 2019-04-24 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Brainlab Ag |
| Distribution | US |
Product
Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine & Trauma Navigation System is intended as an intraoperative image-guided localization system to enable minimally invasive surgery.
Reason
In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment.
Identifiers
| code_info | Version 1.0.0.249, GTIN 04056481140328 |
|---|