FDA Device recall Z-1081-2015

VisiPlugST for the Lacrimal Efficiency Test

On 2015-02-18 the FDA classified a Class III recall of VisiPlugST for the Lacrimal Efficiency Test by Lacrimedics, citing moisture level may cause the VisiPlugST (or CollaSyn Plugs) to break apart, or crumble upon contact. The firm's data also suggests that a 4 year shelf life may not be sustained.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1081-2015
ClassificationClass III
Statusterminated
Initiated2014-11-26
Published2015-02-18
Last updated2026-09-13 11:53 UTC
CompanyLacrimedics
DistributionUS

Product

VisiPlugST for the Lacrimal Efficiency Test. It is also sold in the EU as CollaSyn Plugs. Model number 1813, 1814, 1815, 1903, 1904, and 1905. This device is cylindrical in shape, approximately 1.75-2.00mm in length, and available in three sizes: 0.3mm, 0.4mm and 0.5mm. Plugs are place in the upper and/or lower horizontal canaliculus. Two plugs are placed into a foam carrier and nitrogen sealed inside a foil pouch. The foil pouch is sent to sterilizer for gamma sterilization. Two sterile pouches are packaged in a box with an IFU.

Reason

Moisture level may cause the VisiPlugST (or CollaSyn Plugs) to break apart, or crumble upon contact. The firm's data also suggests that a 4 year shelf life may not be sustained.

Identifiers

code_info
LOT numbers: 072314-2328; 072314-2329; 072314-2330; 081314-2365; 081314-2366; 081314-2367; 091814-2345;…LOT numbers: 072314-2328; 072314-2329; 072314-2330; 081314-2365; 081314-2366; 081314-2367; 091814-2345; 091814-2350; 100614-2344; EXPIRATION DATES: 2018-05-15; 2018-07-23; 2018-08-13; 2018-08-14; 2018-09-14; 2018-09-14; 2018-09-25; 2018-10-06;