RecallRadar

FDA Device recall Z-1079-2014

Trabecular Metal" Reverse Shoulder System

On 2014-03-05 the FDA classified a Class II recall of Trabecular Metal" Reverse Shoulder System by Zimmer, citing as a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1079-2014
ClassificationClass II
Statusterminated
Initiated2013-10-17
Published2014-03-05
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

Trabecular Metal" Reverse Shoulder System

Reason

As a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additional instructions to users on how to properly assemble the Glenosphere to the Baseplate.

Identifiers

code_info
61025613, 61009830, 61038635, 61031385, 61021507, 61021506, 61016907, 61009834, 61006493, 61006492, 60999239, 60993837,…61025613, 61009830, 61038635, 61031385, 61021507, 61021506, 61016907, 61009834, 61006493, 61006492, 60999239, 60993837, 60993836, 60987295, 60982839, 60982838, 60970974, 60970973, 60950253, 60964425, 60946627, 60970946, 60964426, 60943695, 60940448, 60950251, 60946624, 60940449, 60927405, 60934492, 60900542, 60927406, 60919029, 60914617, 60919028, 60900543, 60893180, 60893181, 60906211, 60887745, 60887744, 60875337, 60875336, 60868040, 60873954, 60868041, 60867386, 60867385, 60856882, 60856880, 60855532, 60852815, 60855529, 60846158, 60846157, 60840619, 60834839, 60806512, 60824952, 60820211, 60814537, 60785523, 60795705, 60792083, 60761203, 60752933, 60781096, 60776583, 60765951, 60768804, 60757206, 60747093, 60741590, 60747092, 60741589, 60738183, 60727960, 60738184, 60730072, 60730073, 60723302, 60723277, 60718989, 60712568, 60683321, 60675599, 60669310, 60674468, 105533, 60605155, 105588, 60577323, 60580942, 60572796, 60564704, 60572798, 60568536, 60568535, 60578677, 60568468, 6056

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1079-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.