FDA Device recall Z-1077-2024

BEQ-HLS 7050 USA, HLS Set Advanced 7.0, Product Code/Part Number 70106.9078

On 2024-02-21 the FDA classified a Class II recall of BEQ-HLS 7050 USA, HLS Set Advanced 7.0, Product Code/Part Number 70106.9078 by Maquet Medical Systems Usa, citing the Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1077-2024
ClassificationClass II
Statusongoing
Initiated2024-01-16
Published2024-02-21
Last updated2026-09-13 11:57 UTC
CompanyMaquet Medical Systems Usa
DistributionUS

Product

BEQ-HLS 7050 USA, HLS Set Advanced 7.0, Product Code/Part Number 70106.9078. Used with the CARDIOHELP heart-lung machine.

Reason

The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection.

Identifiers

code_info
UDI-DI 04058863080383 Lots 3000280906, 3000280907, 3000280908, 3000280909, 3000282253, 3000282254, 3000282255, 30002911…UDI-DI 04058863080383 Lots 3000280906, 3000280907, 3000280908, 3000280909, 3000282253, 3000282254, 3000282255, 3000291127, 3000300688, 3000314506, 3000316206, 3000316600, 3000317531, 3000317532, 3000318850, 3000318851, 3000318852, 3000318853, 3000319748, 3000319749, 3000319750, 3000319751, 3000319778, 3000319781, 3000323082, 3000323084, 3000323085, 3000324495, 3000324496, 3000324497, 3000324499, 3000325557, 3000325558, 3000325559, 3000327120, 3000327121, 3000328404, 3000328405, 3000334413, 3000334414, 3000336487 ***Added 3/21/24*** 3000330091, 3000330092, 3000330094, 3000330095, 3000330096, 3000333095, 3000333096, 3000333097, 3000334409, 3000334410, 3000336488, 3000337834, 3000342904, 3000342905, 3000342906, 3000342907, 3000342908, 3000344446, 3000344447, 3000344471, 3000344472, 3000344474, 3000346048, 3000346049, 3000346050, 3000346051, 3000346052, 3000348543, 3000348546, 3000348552, 3000348553, 3000349793, 3000349795, 3000349796, 3000349797, 3000352614, 3000352615, 3000352616, 30003

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1077-2024%22