RecallRadar

FDA Device recall Z-1076-2024

Model No. BEQ-HLS 5050 USA, HLS Set Advanced 5.0, Product Code/Part Number 70106.9077

On 2024-02-21 the FDA classified a Class II recall of Model No. BEQ-HLS 5050 USA, HLS Set Advanced 5.0, Product Code/Part Number 70106.9077 by Maquet Medical Systems Usa, citing the Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1076-2024
ClassificationClass II
Statusongoing
Initiated2024-01-16
Published2024-02-21
Last updated2026-09-13 11:57 UTC
CompanyMaquet Medical Systems Usa
DistributionUS

Product

Model No. BEQ-HLS 5050 USA, HLS Set Advanced 5.0, Product Code/Part Number 70106.9077. Used with the CARDIOHELP heart-lung machine.

Reason

The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection.

Identifiers

code_infoUDI-DI 04058863076355 Lots 3000301222, 3000319782, 3000327602 ***Added 3/21/24*** 3000333098, 3000352360, 3000357174

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1076-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.