FDA Device recall Z-1075-2024

Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184

On 2024-02-21 the FDA classified a Class II recall of Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184 by Microbiologics, citing the negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the neg….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1075-2024
ClassificationClass II
Statusongoing
Initiated2024-01-08
Published2024-02-21
Last updated2026-09-13 11:57 UTC
CompanyMicrobiologics
DistributionUS

Product

Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.

Reason

The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.

Identifiers

code_info
Lot #8184-68, Exp. 5/31/2025, UDI-DI 70845357041448. UPDATE: The recall was expanded on 1/23/2024 to include the foll…Lot #8184-68, Exp. 5/31/2025, UDI-DI 70845357041448. UPDATE: The recall was expanded on 1/23/2024 to include the following additional lot numbers: 8184-65, exp. 3/31/2025; 8184-66, exp. 4/30/2025; 8184-67, exp. 5/31/2025; 8184-69, exp. 6/30/2025; 8184-70, exp. 8/31/2025; and 8184-71, exp. 8/31/2025.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.