FDA Device recall Z-1074-2013

All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System Kits - Meter

On 2013-04-24 the FDA classified a Class I recall of All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System Kits - Meter by Lifescan, citing the Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying EXTREME HIGH GLUCOSE.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1074-2013
ClassificationClass I
Statusterminated
Initiated2013-03-11
Published2013-04-24
Last updated2026-09-13 11:53 UTC
CompanyLifescan
DistributionUS

Product

All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System Kits - Meter; Verio IQ Starter kits - Meter; Verio IQ Warranty Meter. Product Usage: The OneTouch¿ VerioIQ Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The system is intended to be used by a single patient and should not be used for testing multiple patients. The OneTouch¿ VerioIQ Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.

Reason

The Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying EXTREME HIGH GLUCOSE.

Identifiers

code_info
Part numbers: Verio IQ Systems Kits: 022 267-01, 022 267-01, 022 267-03; Verio IQ Starter Kits: 022 268 01, 022 268-…Part numbers: Verio IQ Systems Kits: 022 267-01, 022 267-01, 022 267-03; Verio IQ Starter Kits: 022 268 01, 022 268-02; Verio IQ Warranty Kits: 022 275 01, 022 275 02. All lots and serial numbers