RecallRadar

FDA Device recall Z-1071-2021

CRESCENT SNARE, model no

On 2021-02-17 the FDA classified a Class II recall of CRESCENT SNARE, model no by Aomori Olympus, citing inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1071-2021
ClassificationClass II
Statusterminated
Initiated2021-01-04
Published2021-02-17
Last updated2026-09-13 11:56 UTC
CompanyAomori Olympus
DistributionUS

Product

CRESCENT SNARE, model no. SD-221U-25 - Product Usage: Used with endoscopes to electrosurgically resect tissue within the digestive tract.

Reason

Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.

Identifiers

code_info01K, 02K, 03K, 04K, 05K, 07K, 08K, 09K, 0XK, 92K, 93K, 94K, 95K, 96K, 97K, 98K, 99K, 9XK, 9YK, 9ZK

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1071-2021%22