FDA Device recall Z-1070-2019
ICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 152 cm (60") Tubing, Disposable Transducer, 03 ml S…
- FDA Device
- Class II
- terminated
- Published 2019-04-03
On 2019-04-03 the FDA classified a Class II recall of ICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 152 cm (60") Tubing, Disposable Transducer, 03 ml S… by Icu Medical, citing reports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1070-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-11-27 |
| Published | 2019-04-03 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Icu Medical |
| Distribution | CA, GA, NC, NY, PA, TX, WI |
Product
ICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 152 cm (60") Tubing, Disposable Transducer, 03 ml Squeeze Flush Device, Macrodrip (Pole Mount), REF: 01C-42640-06, Sterile EO
Reason
Reports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay in therapy, and in rare circumstances, pressure tubing separations could lead to blood loss, air embolism, or contamination of the fluid path.
Identifiers
| code_info | Lot/UDI: 3514084/ (01)0 0840619 03747 5 (17) 201001 (30) 01 (10) 3514084; 3550279/ (01)0 0840619 03747 5 (17) 201001 (30) 01 (10) 3550279 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1070-2019%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.