Human Subclass Kit for use on the Beckman IMMAGE"/IMMAGE"800 Analyser Product Code LK009.IM This kit is intended for qualifying human IgG s…

On 2017-02-01 the FDA classified a Class II recall of Human Subclass Kit for use on the Beckman IMMAGE"/IMMAGE"800 Analyser Product Code LK009.IM This kit is intended for qualifying human IgG s… by The Binding Site Group, citing customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1067-2017
ClassificationClass II
Statusterminated
Initiated2014-09-11
Published2017-02-01
Last updated2026-09-13 11:54 UTC
CompanyThe Binding Site Group
DistributionCA, FL, MI, MN, OH, TN

Product

Human Subclass Kit for use on the Beckman IMMAGE"/IMMAGE"800 Analyser Product Code LK009.IM This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents

Reason

Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.

Identifiers

code_infoAll lots.