FDA Device recall Z-1067-2016
iChem VELOCITY Automated Urine Chemistry, Catalog No. 800-3049, 800-3050, 800-3079, 800-3080, 800-3530, 800-7167, 800-7190, 800-7713, 800-7…
- FDA Device
- Class II
- terminated
- Published 2016-03-16
On 2016-03-16 the FDA classified a Class II recall of iChem VELOCITY Automated Urine Chemistry, Catalog No. 800-3049, 800-3050, 800-3079, 800-3080, 800-3530, 800-7167, 800-7190, 800-7713, 800-7… by Beckman Coulter, citing iris International noted loose or missing analyte pads found in several locations (strip vial, strip provider module, strip conveyor system or waste container) that are undetectab….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1067-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-01-13 |
| Published | 2016-03-16 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Beckman Coulter |
| Distribution | PR |
Product
iChem VELOCITY Automated Urine Chemistry, Catalog No. 800-3049, 800-3050, 800-3079, 800-3080, 800-3530, 800-7167, 800-7190, 800-7713, 800-7714, 800-7719, 800-7720 iChem VELOCITY Urine Chemistry Strips, Catalog No. 800-7212 & 800-7212-001
Reason
Iris International noted loose or missing analyte pads found in several locations (strip vial, strip provider module, strip conveyor system or waste container) that are undetectable by the iChemVELOCITY System.
Identifiers
| code_info | all serial numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1067-2016%22