FDA Device recall Z-1066-2017
Apollo EZ: Model: 9784152036 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures
- FDA Device
- Class II
- ongoing
- Published 2017-02-01
On 2017-02-01 the FDA classified a Class II recall of Apollo EZ: Model: 9784152036 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures by Villa Radiology Systems, citing the analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to acce….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1066-2017 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2016-02-19 |
| Published | 2017-02-01 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Villa Radiology Systems |
| Distribution | AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, OH, PA, PR, TN, TX, UT, WI |
Product
Apollo EZ: Model: 9784152036 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.
Reason
The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
Identifiers
| code_info | 13101261 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1066-2017%22