FDA Device recall Z-1064-2018
6.5mm x 125mm x 16mm Cannulated Screw, Item Number: 110009948 (UDI: (01)00887868111327 (17)270421 (10)120910) Product Usage: The 6.5/8.0mm…
- FDA Device
- Class II
- terminated
- Published 2018-03-21
On 2018-03-21 the FDA classified a Class II recall of 6.5mm x 125mm x 16mm Cannulated Screw, Item Number: 110009948 (UDI: (01)00887868111327 (17)270421 (10)120910) Product Usage: The 6.5/8.0mm… by Zimmer Biomet, citing one lot of 3.5mm x 24mm Cortical Locking screws is labeled as 6.5mm x 125mm x 16mm Cannulated screws.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1064-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-12-07 |
| Published | 2018-03-21 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Zimmer Biomet |
| Distribution | n/a |
Product
6.5mm x 125mm x 16mm Cannulated Screw, Item Number: 110009948 (UDI: (01)00887868111327 (17)270421 (10)120910) Product Usage: The 6.5/8.0mm Cannulated Screw System is part of a series of cannulated screws ranging from 4.0 to 8.0mm. Used in Slipped capital femoral epiphysis, Pediatric femoral neck fractures, Tibial plateau fractures, SI joint disruptions, Intercondylar femur fractures, Subtalar arthrodesis, and Fixation of pelvis and iliosacral joint.
Reason
One lot of 3.5mm x 24mm Cortical Locking screws is labeled as 6.5mm x 125mm x 16mm Cannulated screws.
Identifiers
| code_info | Lot: 120910 |
|---|
Status history