AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2

On 2024-03-06 the FDA classified a Class I recall of AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2 by Vyaire Medical, citing risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ve….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1063-2024
ClassificationClass I
Statusongoing
Initiated2024-01-10
Published2024-03-06
Last updated2026-09-13 11:57 UTC
CompanyVyaire Medical
DistributionUS

Product

AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2

Reason

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

Identifiers

code_infoUDI/DI Case: 50190752114164, Each: 10190752114166; All manufacturing dates of 2017 and prior, and any products without a manufacturing date.