FDA Device recall Z-1063-2024
AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2
- FDA Device
- Class I
- ongoing
- Published 2024-03-06
On 2024-03-06 the FDA classified a Class I recall of AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2 by Vyaire Medical, citing risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ve….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1063-2024 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2024-01-10 |
| Published | 2024-03-06 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Vyaire Medical |
| Distribution | US |
Product
AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2
Reason
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
Identifiers
| code_info | UDI/DI Case: 50190752114164, Each: 10190752114166; All manufacturing dates of 2017 and prior, and any products without a manufacturing date. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1063-2024%22