FDA Device recall Z-1053-2023
GE Nuclear Medicine 600/800 series systems, specifically: a) 870, b) 870 DR 3/8 inch NM GANTRY WITH NEW ROTOR - MEA, c) 870 DR / Dismount,…
- FDA Device
- Class I
- ongoing
- Published 2023-02-15
On 2023-02-15 the FDA classified a Class I recall of GE Nuclear Medicine 600/800 series systems, specifically: a) 870, b) 870 DR 3/8 inch NM GANTRY WITH NEW ROTOR - MEA, c) 870 DR / Dismount,… by Ge Medical Systems Israel Functional Imaging, citing A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1053-2023 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2022-12-19 |
| Published | 2023-02-15 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Ge Medical Systems Israel Functional Imaging |
| Distribution | US |
Product
GE Nuclear Medicine 600/800 series systems, specifically: a) 870, b) 870 DR 3/8 inch NM GANTRY WITH NEW ROTOR - MEA, c) 870 DR / Dismount, d) 870 DR / Dismount / 60HZ, e) 870 DR 3/8 inch WSO, f) 870 DR 3/8 inch NM Gantry, g) 870 DR 3/8 inch, h) 870 DR 3/8 inch CH, i) 870 DR 5/8 inch NM GANTRY, j) 870 DR BJ / Dismount, k) 870 DR CyberDefense 3/8 inch NM Gantry, l) 870 DR NM GANTRY FINAL - NEW ROTOR, m) 870 DR China, n) D670 NM GANTRY WITH NEW ROTOR, o) D670 DR 3/8 inch 16 SL, p) D670 DR 3/8 inch MEA, q) D670 DR 3/8 inch, r) D670 DR 3/8 inch 16SL, s) D670 DR 3/8 inch 16SL NR, t) D670 DR 3/8 inch BJ, u) D670 DR 3/8 inch MEA, v) D670 DR NM Gantry 3/8 inch, w) D670 DR NM Gantry 5/8 inch, x) D670 DR with new rotor, y) D670 ES NM Gantry 3/8 inch, z) D670 ES NM/CT 3/8 inch O540+ NM Gantry, aa) D670 ES NM/CT DR 3/8 inch, bb) D670 GANTRY, cc) D670 PRO NM Gantry 3/8 inch, dd) D670 PRO NM GANTRY 3/8 inch, ee) DISC NM/CT670 DR 3/8 inch, ff) DISC NM/CT870 DR 3/8 inch, gg) DISC NM/CT870 DR 3/8 inch, hh) DISC NM/CT870 DR 5/8 inch, ii) DISCOVERY 670 (TANDEM), jj) DISCOVERY NM 670, kk) DISCOVERY NM/CT 670 3/8 inch PRO, ll) DISCOVERY NM/CT 670 3/8 inch PRO NR, mm) HP Z420 WS Config 1 Hardware Specs Take 3 - E5-1620-8G-1TB-NVS, nn) NM/CT 870 DR, oo) NM/CT 870 DR 3/8 inch CYBER DEFENCE, pp) NM/CT 870 DR 3/8 inch WSO, qq) NM/CT 870 DR O540 3/8 inch, rr) NM/CT 870 DR O540 3/8 inch BJ, ss) NM/CT 870 DR O540 5/8 inch, tt) NM/CT 870 ES 3/8 inch, uu) NM/CT 870DR China for China, vv) NM/CT870 DR 3/8 inch O540, ww) NM/CT D870 DR O540 3/8 inch, xx) TANDEM DISCOVERY 670 INT; Emission Computed Tomography System
Reason
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
Identifiers
| code_info | a) 870: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3DG2200004, NT3DG2200009, NT3DG2200018…a) 870: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3DG2200004, NT3DG2200009, NT3DG2200018, NT3DG2200002; b) 870 DR 3/8" NM GANTRY WITH NEW ROTOR - MEA: Catalog number 5376204-70-60-1, UDI/DI Not Applicable, Serial Numbers: NT3DG2100018BH; c) 870 DR / Dismount: Catalog number To be provided, UDI/DI To be provided, Serial Numbers: NT3DG2100006; d) 870 DR / Dismount / 60HZ: Catalog number To be provided, UDI/DI To be provided, Serial Numbers: NT3DG2000016BH; e) 870 DR 3/8" WSO: Catalog number H3906EM, UDI/DI Not Applicable, Serial Numbers: 870B64224, 870Z64148; f) 870 DR 3/8" NM Gantry: Catalog number H3100AS, UDI/DI Not Applicable, Serial Numbers: 870Y64064, 870Y64040, 870A64172, 870Z64145, 870A64161, 870Z64111, 870Z64126, 870Y64081, 870A64210, 870X64003, 870A64180, 870Y64047, 870Y64044, 870Y64079, 870Y64050, 870Z64130, 870X64011, 870Y64095, 870Y64037, 870Y64074, 870Y64067, 870Y64051, 870A64162, 870X64024, 870A64163, 870A64165, 870X64002, 870Y640 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1053-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.