RecallRadar

FDA Device recall Z-1047-2025

regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator

On 2025-02-12 the FDA classified a Class I recall of regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator by Roi Cps, citing there is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1047-2025
ClassificationClass I
Statusongoing
Initiated2024-12-26
Published2025-02-12
Last updated2026-09-13 11:57 UTC
CompanyRoi Cps
DistributionMO

Product

regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.

Reason

There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..

Identifiers

code_info
Lot number 104006, Exp. 9/30/2026, UDI (01)10194717119074(17)260930(10)104006; and Lot number 103486, Exp. 8/31/2026, U…Lot number 104006, Exp. 9/30/2026, UDI (01)10194717119074(17)260930(10)104006; and Lot number 103486, Exp. 8/31/2026, UDI (01)10194717119074(17)260831(10)103486.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1047-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.