FDA Device recall Z-1046-2017
Merge Cardio software
- FDA Device
- Class II
- terminated
- Published 2017-01-25
On 2017-01-25 the FDA classified a Class II recall of Merge Cardio software by Merge Healthcare, citing if a reader selects Multi-study review prior to the original study completing its loading process and then immediately returns to image review of the original study, it is possibl….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1046-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-03-28 |
| Published | 2017-01-25 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Merge Healthcare |
| Distribution | US |
Product
Merge Cardio software. Product Usage: Merge Cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information including images, Hemodynamic studies and reports, measurements (via import from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.
Reason
If a reader selects Multi-study review prior to the original study completing its loading process and then immediately returns to image review of the original study, it is possible that not all current images will be present for review, which could result in incorrect treatment/diagnosis.
Identifiers
| code_info | Versions: 9.0, 9.0.1, 9.0.2, 9.0.3, 9.0.4, 9.0.5, 9.0.6, 9.0.7, 9.0.8, 9.0.9, 10.0, 10.0.1, 10.1, 10.1.1, and 10.1.2 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1046-2017%22