RecallRadar

FDA Device recall Z-1041-2022

Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient…

On 2022-05-18 the FDA classified a Class II recall of Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient… by Philips North America, citing when a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to N….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1041-2022
ClassificationClass II
Statusterminated
Initiated2022-02-22
Published2022-05-18
Last updated2026-09-13 11:56 UTC
CompanyPhilips North America
DistributionFL, ID, NJ, OK, SC, TX

Product

Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient hemodynamic data in the Philips Interventional Hemodynamic System.

Reason

When a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, mmHg nerve damage or bone deformity for a neonate.

Identifiers

code_info
UDI: B-00884838084544; System Serial Numbers: 396 411 225 351 51 46 33 189 132 19 51 924 644 645 646 376 384 487 538 8…UDI: B-00884838084544; System Serial Numbers: 396 411 225 351 51 46 33 189 132 19 51 924 644 645 646 376 384 487 538 886 14 15 643 1194 220 1238 947 949 950 951 89 657 169 168 413 414 624 565 1060 393 512 278 818 362 123 802 395 576 932 291 834 539 605 161 970 686 761 229 230

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1041-2022%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.