RecallRadar

FDA Device recall Z-1040-2024

BD BBL Sensi Disc Tetracycline 30 ug,used for semi-quantitative in vitro susceptibility testing

On 2024-02-14 the FDA classified a Class II recall of BD BBL Sensi Disc Tetracycline 30 ug,used for semi-quantitative in vitro susceptibility testing by Becton Dickinson And, citing there is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed resul….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1040-2024
ClassificationClass II
Statusongoing
Initiated2024-01-08
Published2024-02-14
Last updated2026-09-13 11:57 UTC
CompanyBecton Dickinson And
DistributionUS

Product

BD BBL Sensi Disc Tetracycline 30 ug,used for semi-quantitative in vitro susceptibility testing. Catalog Number 230998 and 231344¿

Reason

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.

Identifiers

code_info
Catalog No. 230998 UDI-DI 00382902309987 Lots 1242880 1277918 1301570 2025088 2056233 2118569 2276681 3094187 3212760 3…Catalog No. 230998 UDI-DI 00382902309987 Lots 1242880 1277918 1301570 2025088 2056233 2118569 2276681 3094187 3212760 3263267 9009914 9046895 0065366 0178210; Catalog No. 231344¿ UDI-DI 30382902313442 Lots 3010027 3062263 3094187 3158029 3212760 3263267 3275644 0065366 0085202 0230453 0274792 0335906 1095603 1148539 1180271 1242880 1277918 1301570 2025088 2056233 2089874 2118569 2276681 8355562 9009914 9046895 9130797 9239639 9263061

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1040-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.