FDA Device recall Z-1039-2024

BD BBL Sensi Disc Sulfamethoxazole with Trimethoprim 23.75/1.25 ug, used for semi-quantitative in vitro susceptibility testing

On 2024-02-14 the FDA classified a Class II recall of BD BBL Sensi Disc Sulfamethoxazole with Trimethoprim 23.75/1.25 ug, used for semi-quantitative in vitro susceptibility testing by Becton Dickinson And, citing there is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed resul….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1039-2024
ClassificationClass II
Statusongoing
Initiated2024-01-08
Published2024-02-14
Last updated2026-09-13 11:57 UTC
CompanyBecton Dickinson And
DistributionUS

Product

BD BBL Sensi Disc Sulfamethoxazole with Trimethoprim 23.75/1.25 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231536 and 231539¿

Reason

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.

Identifiers

code_info
Catalog No. 231536 UDI-DI 00382902315360 Lots 1060689 1116935 1210905 1242939 1306103 1334461 2004540 2025116 2056456 2…Catalog No. 231536 UDI-DI 00382902315360 Lots 1060689 1116935 1210905 1242939 1306103 1334461 2004540 2025116 2056456 2108378 2146453 2301307 3003869 3062089 3094151 3118079 3152968 3248444 0309595 0343394; Catalog No. 231539¿ UDI-DI 30382902315392 Lots 2339393 3003869 3062089 3094151 3118079 3152968 3214768 3248444 3282335 0309595 0343394 1060689 1116935 1148543 1210905 1242939 1306103 1334461 2004540 2025116 2056456 2108378 2146453 2277321 2301307

Status history