FDA Device recall Z-1038-2024
BD BBL Sensi Disc Rifampin- 5 ug, used for semi-quantitative in vitro susceptibility testing
- FDA Device
- Class II
- ongoing
- Published 2024-02-14
On 2024-02-14 the FDA classified a Class II recall of BD BBL Sensi Disc Rifampin- 5 ug, used for semi-quantitative in vitro susceptibility testing by Becton Dickinson And, citing there is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed resul….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1038-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-01-08 |
| Published | 2024-02-14 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Becton Dickinson And |
| Distribution | US |
Product
BD BBL Sensi Disc Rifampin- 5 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231541 and 231544¿
Reason
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
Identifiers
| code_info | Catalog No. 231541 UDI-DI 00382902315414 Lots 1278110 1334479 2004554 2025125 2056466 2242510 3214931; Catalog No. 231…Catalog No. 231541 UDI-DI 00382902315414 Lots 1278110 1334479 2004554 2025125 2056466 2242510 3214931; Catalog No. 231544¿ UDI-DI 30382902315446 Lots 1116055 1211489 2339407 3030278 3158059 3214931 0343413 1032232 1088790 1148547 1278110 1334479 2004554 2025125 2056466 2090443 2242510 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1038-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.