FDA Device recall Z-1037-2024

BD BBL Sensi Disc Piperacillin/Tazobactam- 100/10 ug, used for semi-quantitative in vitro susceptibility testing

On 2024-02-14 the FDA classified a Class II recall of BD BBL Sensi Disc Piperacillin/Tazobactam- 100/10 ug, used for semi-quantitative in vitro susceptibility testing by Becton Dickinson And, citing there is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed resul….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1037-2024
ClassificationClass II
Statusongoing
Initiated2024-01-08
Published2024-02-14
Last updated2026-09-13 11:57 UTC
CompanyBecton Dickinson And
DistributionUS

Product

BD BBL Sensi Disc Piperacillin/Tazobactam- 100/10 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231691 and 231692¿

Reason

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.

Identifiers

code_info
Catalog No. 231691 UDI-DI 00382902316916 Lots 1301719 2026369 2056587 2119595 3003951 3062107 3118199 3275730; Catalog…Catalog No. 231691 UDI-DI 00382902316916 Lots 1301719 2026369 2056587 2119595 3003951 3062107 3118199 3275730; Catalog No. 231692¿ UDI-DI 30382902316924 Lots 3003951 3184117 3223408 3275730 1301719 2026369 2056587 2119595 2147716 2244463 3062107 3118199