RecallRadar

FDA Device recall Z-1035-2016

GlideScope Titanium Single-Use Video Laryngoscope

On 2016-03-16 the FDA classified a Class I recall of GlideScope Titanium Single-Use Video Laryngoscope by Verathon, citing the firm has become aware of the potential for disruption (flickering) in the video laryngoscopy image when GlideScope Titanium SU laryngoscope blades are in use. Video "flickerin….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1035-2016
ClassificationClass I
Statusterminated
Initiated2016-01-29
Published2016-03-16
Last updated2026-09-13 11:54 UTC
CompanyVerathon
DistributionUS

Product

GlideScope Titanium Single-Use Video Laryngoscope. It is a single-use video laryngoscope system (GlideScope Titanium System) consists of a sterile-packaged, single-use video laryngoscope, multi-use Smart Cable and portable GlideScope video monitor. The system incorporates a high-resolution, full-color digital camera and monitor for real-time viewing and recording. Affected Product Names and Part Numbers: -LoPro S3: 0574-0130 (single blade) or 0270-0769 (box of 10) -LoPro S4: 0574-0131 (single blade) or 0270-0770 (box of 10) -MAC S3: 0574-0132 (single blade) or 0270-0771 (box of 10) -MAC S4: 0574-0133 (single blade) or 0270-0772 (box of 10) The GlideScope Video Laryngoscopes are intended for use by qualified medical professionals to obtain a clear, unobstructed view of the vocal cords for medical procedures.

Reason

The firm has become aware of the potential for disruption (flickering) in the video laryngoscopy image when GlideScope Titanium SU laryngoscope blades are in use. Video "flickering" appears as the intermittent break-up of the on-screen video image, appearing as either distorted horizontal or vertical bars of displaced video signal.

Identifiers

code_infoLOT Number ranges: -LoPro S3: 081814 - 093015; -LoPro S4: 081114 - 090315; -MAC S3: 080814 - 101315; -MAC S4: 022514 - 082115;

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1035-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.