FDA Device recall Z-1035-2013
UniCel DxC 600i Synchron Access Clinical System
- FDA Device
- Class II
- terminated
- Published 2013-04-10
On 2013-04-10 the FDA classified a Class II recall of UniCel DxC 600i Synchron Access Clinical System by Beckman Coulter, citing beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1035-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-02-12 |
| Published | 2013-04-10 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Beckman Coulter |
| Distribution | US |
Product
UniCel DxC 600i Synchron Access Clinical System; P/N 107405S Product Usage: The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
Reason
Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the light shield can fall into the reaction vessels (RVs) and present a risk of erroneous test results.
Identifiers
| code_info | Part # A25640, A25638, A25656, A25637; All Serial Numbers |
|---|