Animas Vibe Insulin Infusion Pump

On 2015-02-11 the FDA classified a Class II recall of Animas Vibe Insulin Infusion Pump by Animas, citing calibration factors in the pump overwritten during a programming step. The force sensor could send a lower signal value to the pump processor, with loss of prime warnings, occlusi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1034-2015
ClassificationClass II
Statusterminated
Initiated2011-09-06
Published2015-02-11
Last updated2026-09-13 11:53 UTC
CompanyAnimas
Distributionn/a

Product

Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature.

Reason

Calibration factors in the pump overwritten during a programming step. The force sensor could send a lower signal value to the pump processor, with loss of prime warnings, occlusion alarms and the pump unable to detect a cartridge during the prime sequence. Field action initiated 8/29/2011.

Identifiers

code_info
Model Number(s): 100515-63 100510-63 100514-63 100512-63 100511-63 101200-03 100201-03 101202-03 101204-03…Model Number(s): 100515-63 100510-63 100514-63 100512-63 100511-63 101200-03 100201-03 101202-03 101204-03 101205-03 101200-53 101202-53 101205-53 101200-02 101200-57 101201-57 101202-57 101204-57 101205-57 101200-63 101201-63 101202-63 101204-63 101205-63 101206-63