RecallRadar

FDA Device recall Z-1034-2014

6 DEGREE COCR FEM HEAD 36

On 2014-02-26 the FDA classified a Class II recall of 6 DEGREE COCR FEM HEAD 36 by Zimmer, citing zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was deter….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1034-2014
ClassificationClass II
Statusterminated
Initiated2013-10-11
Published2014-02-26
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

6 DEGREE COCR FEM HEAD 36

Reason

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

Identifiers

code_info
369934 60308845 60508319 60588077 60670902 60701935 60779840 60790101 60810357 60847644 60906703 61077837 6…369934 60308845 60508319 60588077 60670902 60701935 60779840 60790101 60810357 60847644 60906703 61077837 61077838 61191243 61191244 61247015 61304729 61340542 61401735 61485227 61530679 61551614 61555020 61670831 61720512 61774878 61790557 61837158 62025229 62124070 62176438 369935 60339031 60407279 60586160 60658085 60701954 60701967 60779852 60779856 60779863 60806208 60821916 60847645 60900931 61002276 61052833 61077839 61077840 61191246 61304732 61401736 61448090 61502826 61530680 61570271 61618471 61618472 61713710 61734540 61837159 61837160 61896789 62025230 62091051 62176441 62284369 60339030 60392901 60440343 60508338 60656885 60669433 60779864 60779866 60779867 60790102 60806209 60821917 60847646 61002277 61052834 61077841 61144531 61191245 61304736 61401737 61485225 61530678 61551616 61551617 61670830 61713709 61734539 61774872 61779845 61790554 61896791 61896792 62025231 6202

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1034-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.