RecallRadar

FDA Device recall Z-1028-2014

VERSYS HERITAGE FEM STEM

On 2014-02-26 the FDA classified a Class II recall of VERSYS HERITAGE FEM STEM by Zimmer, citing zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was deter….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1028-2014
ClassificationClass II
Statusterminated
Initiated2013-10-11
Published2014-02-26
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

VERSYS HERITAGE FEM STEM

Reason

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

Identifiers

code_info
60888636 60953044 60953045 60990022 60990023 61037925 61101347 61190038 61199991 61216692 61231958 61337105…60888636 60953044 60953045 60990022 60990023 61037925 61101347 61190038 61199991 61216692 61231958 61337105 61337106 61337108 61337109 61380551 61393846 61395480 61395481 61395482 61457789 61488608 61501472 61528642 61607607 61613260 61631108 61636270 61640544 61658343 61727080 61795011 61803866 61830030 61841636 61841637 61859438 61860472 61878469 61880368 61903233 61936199 61980174 62001339 62024623 62055989 62055991 62097061 62106808 62143853 62147329 62147330 62199398 62199399 62213245 62232590 62240777 62257146 62268582 62268583 62281833 62298293 62305944 369871 61282966 61727082 61962092 369672 61189806 61266267 61294351 369898 61293625

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1028-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.