FDA Device recall Z-1027-2014
PRESSFIT FEM ST FX 16 X 1
- FDA Device
- Class II
- terminated
- Published 2014-02-26
On 2014-02-26 the FDA classified a Class II recall of PRESSFIT FEM ST FX 16 X 1 by Zimmer, citing zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was deter….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1027-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-10-11 |
| Published | 2014-02-26 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Zimmer |
| Distribution | US |
Product
PRESSFIT FEM ST FX 16 X 1
Reason
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
Identifiers
| code_info | 60621428 60667768 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1027-2014%22