RecallRadar

FDA Device recall Z-1024-2015

Beckman Coulter, Uric Acid Assay, Catalog No. OSR6X98

On 2015-02-04 the FDA classified a Class II recall of Beckman Coulter, Uric Acid Assay, Catalog No. OSR6X98 by Beckman Coulter, citing N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen overdose, interferes with the assays for these analytes: Cholesterol OS….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1024-2015
ClassificationClass II
Statusterminated
Initiated2014-12-15
Published2015-02-04
Last updated2026-09-13 11:53 UTC
CompanyBeckman Coulter
DistributionUS

Product

Beckman Coulter, Uric Acid Assay, Catalog No. OSR6X98

Reason

N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen overdose, interferes with the assays for these analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. NAC interference may lead to falsely low results.

Identifiers

code_infoAll Lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1024-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.