FDA Device recall Z-1019-2013
Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewire, Models ARCSJ200W, ARCSJ260W, and ARCDJ260W. Sterile using ethylene oxide
- FDA Device
- Class II
- terminated
- Published 2013-04-03
On 2013-04-03 the FDA classified a Class II recall of Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewire, Models ARCSJ200W, ARCSJ260W, and ARCDJ260W. Sterile using ethylene oxide by Medtronic Cardiac Rhythm Disease Management, citing medtronic has identified an issue involving specific lot numbers of the Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewire, Models ARCSJ200W, ARCSJ260W, and ARCDJ260W, whe….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1019-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-03-13 |
| Published | 2013-04-03 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Medtronic Cardiac Rhythm Disease Management |
| Distribution | US |
Product
Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewire, Models ARCSJ200W, ARCSJ260W, and ARCDJ260W. Sterile using ethylene oxide. Product Usage: The Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewires are intended to facilitate catheter placement and exchange during diagnostic or interventional procedures in the aorta, where increased support, distal flexibility, and low surface friction of the guidewire is needed.
Reason
Medtronic has identified an issue involving specific lot numbers of the Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewire, Models ARCSJ200W, ARCSJ260W, and ARCDJ260W, where a change in the manufacturing process has been identified as the cause for potential kinking and breaking at the proximal weld end of the outer spring coil during in-vivo use.
Identifiers
| code_info | Model ARCSJ200W Lot Numbers: GFWC1124, GFWC1506, GFWC2138, GFWC2652, GFWC2653, GFWD0353, GFWD0881, GFWD0882, GFWD2048,…Model ARCSJ200W Lot Numbers: GFWC1124, GFWC1506, GFWC2138, GFWC2652, GFWC2653, GFWD0353, GFWD0881, GFWD0882, GFWD2048, GFWD2049, GFWD2050, GFWD2051, GFWD2052, GFWE0743, GFWE0744, GFWE0745, GFWE0746, GFWE0747, GFWE3710, GFWE4272, GFWE4273, GFWE4274, GFWE4275, GFWF0143, GFWF0144, GFWF0147, GFWF0148, GFWF0149, GFWG0428, GFWG0429, GFWG0430, GFWG0431, GFWG0432, GFWG0433, GFWG0434, GFWH0333, GFWH0334, GFWH0335, GFWH0336, GFWH0337, GFWH0338, GFWI0190, GFWI0191, GFWI0192, GFWI0193, GFWI0194, GFWI0195, GFWJ0027, GFWJ0028, GFWJ0029, GFWJ2454, GFWJ2455, GFWJ2456, GFWJ2457, GFWJ2458, GFWK1910, GFWK1911, GFWL0370, GFWL2571, GFWL2572, GFXA0224. Model ARCSJ260W Lot Numbers: GFWC1501, GFWC1502, GFWC1503, GFWD0883, GFWD0884, GFWD3256, GFWE0752, GFWE0753, GFWE3711, GFWF0145, GFWF0150, GFWG0425, GFWG0426, GFWG0427, GFWH0344, GFWH0345, GFWI0201, GFWI0202, GFWI0203, GFWJ0418, GFWJ0419, GFWJ0420, GFWJ2461, GFWJ2462, GFWJ2464, GFWJ2465, GFWL0368, GFWL0369, GFWL0415, GFWL0416, GFXA0219, GFXA0220, GFXA0 |
|---|