RecallRadar

FDA Device recall Z-1017-2015

Covidien Trellis-6 Peripheral Infusion System

On 2015-02-11 the FDA classified a Class I recall of Covidien Trellis-6 Peripheral Infusion System by Covidien, citing A manufacturing error resulted in the risk of incorrect proximal and distal balloon inflation port identification on the units. Units have been identified to have the distal ballo….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1017-2015
ClassificationClass I
Statusterminated
Initiated2014-12-10
Published2015-02-11
Last updated2026-09-13 11:53 UTC
CompanyCovidien
DistributionUS

Product

Covidien Trellis-6 Peripheral Infusion System. Models BVT608010V01, BVT608030V01, BVT612010V01, and BVT612030V01. Sterile EO. Product Usage: Intended for controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.

Reason

A manufacturing error resulted in the risk of incorrect proximal and distal balloon inflation port identification on the units. Units have been identified to have the distal balloon inflation port incorrectly labeled as proximal, and, the proximal balloon port incorrectly labeled as distal.

Identifiers

code_info
Model BVT608010, Lot numbers: 9853174, 9864497, 9890772, 9890931, 9925676, 9932672, 9937308, 9937315, 9940474, 9940614…Model BVT608010, Lot numbers: 9853174, 9864497, 9890772, 9890931, 9925676, 9932672, 9937308, 9937315, 9940474, 9940614, 9976340, 9978233, A001589, A001740. Model BVT608030, Lot Numbers: 9853331, 9887597, 9887695, 9887757, 9932042. Model BVT612010, Lot numbers: 9854121, 9868214, 9868311, 9875472, 9875632, 9876174, 9883295, 9884788, 9884951, 9886196, 9886312, 9932096, 9968511, 9974109, 9974126, 9976472, 9977091, 9978862, A002299, A002423. Model BVT312030, Lot numbers: 9854124, 9892081, 9941213, 9971096, 9971139, 9976506, 9977140, 9978964, 9979548, A003021, A003121, A003651, A007900, A008492, A008603, A008687.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1017-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.