FDA Device recall Z-1017-2015
Covidien Trellis-6 Peripheral Infusion System
- FDA Device
- Class I
- terminated
- Published 2015-02-11
On 2015-02-11 the FDA classified a Class I recall of Covidien Trellis-6 Peripheral Infusion System by Covidien, citing A manufacturing error resulted in the risk of incorrect proximal and distal balloon inflation port identification on the units. Units have been identified to have the distal ballo….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1017-2015 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2014-12-10 |
| Published | 2015-02-11 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Covidien |
| Distribution | US |
Product
Covidien Trellis-6 Peripheral Infusion System. Models BVT608010V01, BVT608030V01, BVT612010V01, and BVT612030V01. Sterile EO. Product Usage: Intended for controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.
Reason
A manufacturing error resulted in the risk of incorrect proximal and distal balloon inflation port identification on the units. Units have been identified to have the distal balloon inflation port incorrectly labeled as proximal, and, the proximal balloon port incorrectly labeled as distal.
Identifiers
| code_info | Model BVT608010, Lot numbers: 9853174, 9864497, 9890772, 9890931, 9925676, 9932672, 9937308, 9937315, 9940474, 9940614…Model BVT608010, Lot numbers: 9853174, 9864497, 9890772, 9890931, 9925676, 9932672, 9937308, 9937315, 9940474, 9940614, 9976340, 9978233, A001589, A001740. Model BVT608030, Lot Numbers: 9853331, 9887597, 9887695, 9887757, 9932042. Model BVT612010, Lot numbers: 9854121, 9868214, 9868311, 9875472, 9875632, 9876174, 9883295, 9884788, 9884951, 9886196, 9886312, 9932096, 9968511, 9974109, 9974126, 9976472, 9977091, 9978862, A002299, A002423. Model BVT312030, Lot numbers: 9854124, 9892081, 9941213, 9971096, 9971139, 9976506, 9977140, 9978964, 9979548, A003021, A003121, A003651, A007900, A008492, A008603, A008687. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1017-2015%22