Dimension Vista 1500, Analyzer, chemistry Model no. 10444801

On 2019-03-27 the FDA classified a Class II recall of Dimension Vista 1500, Analyzer, chemistry Model no. 10444801 by Siemens Healthcare Diagnostics, citing when the Millipore Water Purification Module Progard Pretreatment Packs are used with Dimension Vista Triglyceride assay, patient and QC triglyceride results may be erroneously el….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1015-2019
ClassificationClass II
Statusterminated
Initiated2018-12-12
Published2019-03-27
Last updated2026-09-13 11:55 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Dimension Vista 1500, Analyzer, chemistry Model no. 10444801

Reason

When the Millipore Water Purification Module Progard Pretreatment Packs are used with Dimension Vista Triglyceride assay, patient and QC triglyceride results may be erroneously elevated as a result of contamination.

Identifiers

code_info
Millipore lot number: F7MA50339 F7MA56608 F7NA62488 F8AA13966 F8AA93567 F8BA20457 F8BA35774 F8DA59251 F8HA64477…Millipore lot number: F7MA50339 F7MA56608 F7NA62488 F8AA13966 F8AA93567 F8BA20457 F8BA35774 F8DA59251 F8HA64477 F8HA79819 F8HA91305 F8JA97876 F8MA06374