The GLOBAL UNITE Platform Shoulder System is intended for cemented or uncemented total shoulder or hemi-shoulder replacement procedures, Pr…

On 2018-03-21 the FDA classified a Class II recall of The GLOBAL UNITE Platform Shoulder System is intended for cemented or uncemented total shoulder or hemi-shoulder replacement procedures, Pr… by Depuy Orthopaedics, citing the screw in specific lots of the GLOBAL UNITE Anatomic Body and GLOBAL UNITE Fracture Body was inverted during assembly to the body, which will cause the humeral stem to sit prou….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1012-2018
ClassificationClass II
Statusterminated
Initiated2017-11-28
Published2018-03-21
Last updated2026-09-13 11:54 UTC
CompanyDepuy Orthopaedics
Distributionn/a

Product

The GLOBAL UNITE Platform Shoulder System is intended for cemented or uncemented total shoulder or hemi-shoulder replacement procedures, Product labeling: LE¿O-1100-30-000 REV D.pdf.pdf, LE¿O-1100-40-000 REV D.pdf.pdf, LE¿O-110030100_REV D.pdf.pdf, LE¿O-110030110_REV D.pdf.pdf, LE¿O-110030120_REV D.pdf.pdf

Reason

The screw in specific lots of the GLOBAL UNITE Anatomic Body and GLOBAL UNITE Fracture Body was inverted during assembly to the body, which will cause the humeral stem to sit proud and may cause surgical delays.

Identifiers

code_infoLot # 8556802, 8556803, 8556804, 8556805, 8583381, 8556910, 8556951, 8605849, 8564487, 8572431, 8610357, 8527674.