RecallRadar

FDA Device recall Z-1010-2018

NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction…

On 2018-03-21 the FDA classified a Class II recall of NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction… by Fresenius Medical Care Renal Therapies Group, citing the label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1010-2018
ClassificationClass II
Statusterminated
Initiated2017-09-29
Published2018-03-21
Last updated2026-09-13 11:54 UTC
CompanyFresenius Medical Care Renal Therapies Group
DistributionCA, FL, KS, LA, TX

Product

NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Reason

The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.

Identifiers

code_infoLot # 17KTAC076; UDI-00840861101696

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1010-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.