FDA Device recall Z-1010-2018
NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction…
- FDA Device
- Class II
- terminated
- Published 2018-03-21
On 2018-03-21 the FDA classified a Class II recall of NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction… by Fresenius Medical Care Renal Therapies Group, citing the label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1010-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-09-29 |
| Published | 2018-03-21 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Fresenius Medical Care Renal Therapies Group |
| Distribution | CA, FL, KS, LA, TX |
Product
NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Reason
The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.
Identifiers
| code_info | Lot # 17KTAC076; UDI-00840861101696 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.