FDA Device recall Z-1007-2024
ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844
- FDA Device
- Class II
- ongoing
- Published 2024-02-14
On 2024-02-14 the FDA classified a Class II recall of ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844 by Globus Medical, citing devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an incr….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1007-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-12-15 |
| Published | 2024-02-14 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Globus Medical |
| Distribution | US |
Product
ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844
Reason
Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.
Identifiers
| code_info | UDI-DI 00889095998498 Lots FSB136AA FSB121AA FSB241AA FSB207AA FSB136AA FSB121AA FSB110AA FSB073AA FSA286CA FSA286BA FS…UDI-DI 00889095998498 Lots FSB136AA FSB121AA FSB241AA FSB207AA FSB136AA FSB121AA FSB110AA FSB073AA FSA286CA FSA286BA FSA286AA FSA263DA FSA263EA FSA263CA FSA263AA FSA263BA FSA160AA FSA160BA FSA105BA FSA105AA FSA053AA FSY319AA FSY279BA FSY279AA FSY272AA FSY230AA FSY098AA FSY098CA FSY098BA FSY048AA FSX093AA FSX205AA FSX141BA FSX141CA FSX141AA FSX093AA FSX063AA FSX041AA FSX041BA FSW142AA FSW094AA |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1007-2024%22