RecallRadar

FDA Device recall Z-0993-2020

Genius 2 Tympanic Thermometer OEM Tympanic Item Code: 303063

On 2020-02-12 the FDA classified a Class II recall of Genius 2 Tympanic Thermometer OEM Tympanic Item Code: 303063 by Cardinal Health 200, citing the frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0993-2020
ClassificationClass II
Statusterminated
Initiated2019-10-28
Published2020-02-12
Last updated2026-09-13 11:55 UTC
CompanyCardinal Health 200
DistributionUS

Product

Genius 2 Tympanic Thermometer OEM Tympanic Item Code: 303063

Reason

The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.

Identifiers

code_infoAll product manufactured after December 10, 2017; Serial Numbers greater than or equal to N17606406

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0993-2020%22