FDA Device recall Z-0987-2015
ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is…
- FDA Device
- Class II
- terminated
- Published 2015-01-21
On 2015-01-21 the FDA classified a Class II recall of ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is… by Siemens Healthcare Diagnostics, citing enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the re….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0987-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-12-16 |
| Published | 2015-01-21 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | WA |
Product
ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.
Reason
Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result could be erroneous and critical and appear to be verified upon repeat.
Identifiers
| code_info | CentraLink" Data Management System V14x Catalog Numbers: 10817209, 10818262 |
|---|