FDA Device recall Z-0982-2014
GE Healthcare, Dash 3000, Dash 4000 and Dash 5000 Patient Monitors, Dash Port2 Docking Station
- FDA Device
- Class II
- terminated
- Published 2014-02-26
On 2014-02-26 the FDA classified a Class II recall of GE Healthcare, Dash 3000, Dash 4000 and Dash 5000 Patient Monitors, Dash Port2 Docking Station by Ge Healthcare, citing GE has become aware of multiple issues affecting the Dash 3000/4000/5000 Patient Monitor and Dash Port 2 Docking Station: DASH: 1. Dash stops communicating on the Unity Network re….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0982-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-11-27 |
| Published | 2014-02-26 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ge Healthcare |
| Distribution | US |
Product
GE Healthcare, Dash 3000, Dash 4000 and Dash 5000 Patient Monitors, Dash Port2 Docking Station.
Reason
GE has become aware of multiple issues affecting the Dash 3000/4000/5000 Patient Monitor and Dash Port 2 Docking Station: DASH: 1. Dash stops communicating on the Unity Network resulting in Loss of Monitoring at the CARESCAPE" CIC Pro Central Station*. 2. Dash displays distorted ECG waveforms in combination monitoring mode. 3. Dash incorrectly becomes Time Master on the Unity Network¿. 4.
Identifiers
| code_info | Parent SN SBG06451608GA SBG06451606GA SBG06451607GA SBG06451638GA SBG06451605GA A4EH1211G A4EH1212G A4EH1251G…Parent SN SBG06451608GA SBG06451606GA SBG06451607GA SBG06451638GA SBG06451605GA A4EH1211G A4EH1212G A4EH1251G A4EH1284G A4EH1286G A4EH1287G A4EH1288G A4EH1196G A4EH1197G A4EH1199G A4EH1203G A4EH1204G A4EH1281G A4EH1192G A4EH1193G A4EH1214G A4EH1227G A4EH1228G A4EH1216GR A4EH1219G A4EH1220GR A4EH1291G A4EH1191G A4EH1229G A4EH1230G A4EH1231G A4EH1240G A4EH1245GR A4EH1275G A4EH1264G A4EH1195G A4EH1246G A4EH1247G A4EH1267G A4EH1268GR A4EH1269G A4EH1271G A4EH1200G A4EH1249G A4EH1250G A4EH1254G A4EH1256G A4EH1257G A4EH1258G A4EH1201G A4EH1202G A4EH1222G A4EH1223G A4EH1205G A4EH1206G A4EH1224G A4EH1225G A4EH1226G A4EH3954GR A4EH1356G A4EH1359G A4EH1342G A4EH1343G A4EH1344G A4EH1345G A4EH1346G A4EH3471G A4EH3472G A4EH3475G A4EH3477G A4EH3478G A4EH3479G A4EH3483G A4EH3487G A4EH1334G A4EH1335G A4EH1336G A4EH1351G A4EH1353G A4EH1355G A4EH3461G A4EH3467G A4EH3468G A4EH1347G A4EH1339G A4EH1174G A4EH1175G A4EH117 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0982-2014%22